This abstract demonstrates that the optimal TE for a single-echo DSC-MRI acquisition is 30-35 ms (without a preload) and 23-30 ms (with a standard full-dose preload) for a high-grade glioma population. Optimal TEs were calculated using error analysis on T2* values quantified during the first-pass of contrast agent using dual-echo DSC, and were shown to be the weighted-average of these values. Furthermore, we demonstrate that the optimal TE depends on tissue type (tumor/healthy tissue/AIF) and preload.
This abstract and the presentation materials are available to members only; a login is required.