Reproducibility of the 1D-ISIS localized ST experiment for hepatic Pi-to-ATP reaction rate measurement at 7T

Ladislav Valkovic 1,2 , Martin Gajdosk 1 , Marek Chmelk 1 , Wolfgang Bogner 1 , Ivan Frollo 2 , Stephan Gruber 1 , Siegfried Trattnig 1 , and Martin Krssk 1,3

1 High Field MR Centre, Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria, 2 Department of Imaging Methods, Institute of Measurement Science, Slovak Academy of Sciences, Bratislava, Slovakia, 3 Division of Endocrinology and Metabolism, Department of Internal Medicine III, Medical University of Vienna, Vienna, Austria

Our aim was to test the reproducibility/reliability of the fast 31 P saturation transfer (ST) in vivo at 7T for localized measurement of Pi-to-ATP reaction in the liver. Test-retest measurements of the hepatic ST experiment and a contamination check of the spectra from the surrounding muscles were performed. Our data show only negligible fraction of the abdominal muscle signal in the liver spectra and the test-retest variation was very low. Thus we can conclude that fast ST measurements in the liver at 7T are highly reproducible.