APIR4EMC: Autocalibrated Parallel Imaging Reconstruction for Extended Multi-Contrast Imaging

Chaoping Zhang1,2, Alexandra Cristobal-Huerta2, Juan Antonio Hernandez-Tamames2, Stefan Klein1,2, and Dirk H.J. Poot1,2

1Departments of Medical Informatics, Erasmus Medical Center, Rotterdam, Netherlands, 2Departments of Radiology and Nuclear Medicine, Erasmus Medical Center, Rotterdam, Netherlands

Multi-contrast images of the same region are routinely acquired in clinical MRI. This abstract proposes a fast reconstruction method for multi-contrast imaging: Autocalibrated Parallel Imaging Reconstruction for Extended Multi-Contrast (APIR4EMC). Unlike conventional parallel imaging (GRAPPA) which reconstructs the image for individual contrast, APIR4EMC additionally includes the inter-contrast signals in the prediction of unsampled positions in the k-space of each contrast and simultaneously reconstructs images for all contrasts. Experiments show significant improvement of APIR4EMC compared to conventional parallel imaging in terms of artifacts and SNR.

This abstract and the presentation materials are available to members only; a login is required.

Join Here