Quantitative Evaluation of Eddy Current Distortion as Part of Quality Assurance Protocol for Multicenter DTI Trial at 3T
Xiaopeng Zhou 1 , Ken Sakaie 1 , Robert Fox 1 , and Mark Lowe 1
1
The Cleveland Clinic, Cleveland, OH, United
States
Eddy current artifact is different for each scanner
model due to different sequence and calibration
algorithm in use. It is beneficial to monitor eddy
current artifact quantitatively and effectively correct
it in a multi-center DTI trial. A quantitative
evaluation of eddy current method as part of QA protocol
was developed and applied to 27 scanners. It is an
effective method to detect scanners with poor eddy
current calibration and can monitor the eddy current
distortion effects at sites longitudinally. This method
also can quantitatively evaluate how effectively an eddy
current correction method can reduce this artifact.