Abstract #0758
Saturation recovery single-shot acquisition (SASHA) for T 1 mapping in the human heart at 7T
Christopher T. Rodgers 1 , Yuehui Tao 1 , Stefan Piechnik 1 , Alexander Liu 1 , Jane Francis 1 , Stefan Neubauer 1 , and Matthew D. Robson 1
1
University of Oxford, Oxford, Oxon, United
Kingdom
We recently reported normal myocardial T1s at 7T, using
the ShMOLLI+IE sequence on a scanner with 16x1kW RF
channels. We now introduce the 7T SASHA variant of the
saturation recovery single-shot acquisition (SASHA)
sequence for use on scanners with only 8x1kW RF. In
phantoms, 7T SASHA T1s were within 6% of IR-SE
reference T1s for readout FAs≤25. In 10 volunteers,
native 7T SASHA T1s in the septum were 193973ms and
999-1674ms after Dotarem contrast. These values are
consistent with our ShMOLLI+IE T1 of 1925 48 ms. 7T
SASHA paves the way for clinical applications of T1
mapping at 7T.