Improvements of RF field Transmission and Detection Sensitivity for 31P MRS with Ultra High Dielectric Constant (uHDC) Material at 7.0 T
Sebastian Rupprecht 1 , Byeong-Yeul Lee 2 , Wei Luo 1 , Xiao-Hong Zhu 2 , Wei Chen 2 , and Qing X Yang 1,3
1
Radiology, Penn State College of Medicine,
Hershey, PA, United States,
2
Center
for Magnetic Resonance Research, Department of
Radiology, University of Minnesota, Minneapolis, MN,
United States,
3
Neurosurgery,
Penn State College of Medicine, Hershey, PA, United
States
High-field X-nuclei MRS for human application at
high/ultrahigh field often suffers from high SAR,
limited SNR and long scanning time. In this report, we
demonstrate that more than 130% SNR gain was achieved
with ultra high dielectric constant (uHDC) materials
incorporated into the RF coil for 31P MRS and CSI at 7T.
Concomitantly, the RF power optimized for acquiring the
spectra was reduce by 43% with the uHDC material. Our
results suggest that uHDC material can be incorporated
into the RF coil for high-field X-nuclear MRS to
significantly improve SNR and reduce transmission power.