Accuracy of Quantitative 3D DCE-MRI Using Variable Flip Angle T1 Mapping, B1 Correction, and the Bookend Method

Brandon Zanette1, Greg O. Cron2, 3, Thanh B. Nguyen2, 3, Mark E. Schweitzer2, 3, Ian G. Cameron2, 3

1Carleton University, Ottawa, Ontario, Canada; 2Ottawa Hospital Research Institute, Ottawa, Ontario, Canada; 3Radiology, University of Ottawa, Ottawa, Ontario, Canada


For quantitative DCE-MRI, Gd concentration is conventionally estimated via a pre-DCE T1 map using variable flip angles (VFA), which allows conversion of DCE signal to concentration. This approach, however, is sensitive to B1 variations. Two techniques may potentially reduce errors: B1 mapping and the Bookend Method (involving a post-DCE T1 map). The results of this study show that, when estimating concentration in tissue for quantitative DCE-MRI, the best results will be achieved using the Bookend Method and accurate T1 maps. B1-corrected VFA does not consistently deliver enough T1 accuracy, indicating that further improvements in T1 mapping techniques are required.