Bo Joseph Peng1, Yibao Wu1,
Jeffrey Walton2, Simon R. Cherry1
1Biomedical Engineering, University of
California, Davis, Davis, CA, United States; 22NMR Facility,
University of California, Davis, Davis, CA, United States
From
single PET module data acquisitions, we further demonstrate the effectiveness
of concentric carbon fiber tubing as an excellent shielding material against
300 MHz RF. We also present a solution to reduce the 81 kHz RF interference
generated by the MR gradient power supply.