Xiaoyu Yang1, Steven Walk1, Paul
Taylor1, Tsinghua Zheng1, Hiroyuki Fujita1,2
1Quality Electrodynamics,
The
MRI trend sees the increasing availability of wider-bore scanners at 1.5T and
3T to accommodate much broader coverage of the patient population. Addressing
the need, an optimized 6-channel ergonomically-designed shoulder coil is
proposed at 1.5T. The coil consists of
3 rows of loop and saddle pairs with flexible flaps for better fitting
of different size shoulder sizes and thereby increasing SNR. Comparison tests
were performed between the proposed flexible coil and a commercially
available 4-channel rigid shoulder coil. The testing and evaluation also
included the performance comparison among various shoulder sizes. The results
show that the proposed one-fits-all coil provides good SNR, depth coverage and uniformity for
the broad range patient population.