Modhurin Banerjee Snyder1,
1GE Healthcare,
In
this work we have created a flexible, modular 32-channel array for
cardio-thoracic imaging that is based on traditional loop elements and Double
Asymmetric Saddle (DAS) pairs. This unique design, dubbed the QASCI (Quad
Asymmetric Saddle for Cardiac Imaging), is an extension of the work done with
the DLAS (Double Loop Asymmetric Saddle) to a cardiothoracic application. The
QASCI system was evaluated via phantom imaging, and demonstrated a nominal
50% improvement in SNR over a larger FOV (34cm by 34 cm) than the 8 channel
cardiac coil, even when evaluated on an element-by-element/channel-by-channel
basis.