Bart Lowie
van de Bank1, Vincent Oltman Boer1, Peter R. Luijten1,
Dennis W.J. Klomp1
1Radiology, University Medical Center Utrecht,
Utrecht, Netherlands
A
dedicated double-tuned double-channel transmit receive surface coil setup is
developed that enables the use of high B1+-fields in multi nuclei MRS of the
human brain at 7T. The available B1+ field of up to 40 T and 100 T for
respectively 1H and 31P allowed the use of short and high bandwidth adiabatic
RF pulses, which are insensitive to the inhomogeneous nature of the
B1+-field. Therefore accurately localized 1H and 31P MR spectra with high
sensitivity could be obtained at 7T.