Daniel Nicolas Splitthoff1, Maxim Zaitsev1
1Dept. of Diagnostic Radiology, Medical
Physics,
With
the SENSE Shimming (SSH) approach a method was introduced recently that
allows for estimating B0 field inhomogeneities based on a reference image and
a series of points on a single free induction decay (FID). In the original
approach the temporal evolution of the FID data is explained by field
inhomogeneities, using linear approximations. Effects caused by relaxation
and those caused by inhomogeneities can therefore not be distinguished and
values can only be given relative to a baseline measurement. We here present
an extension to the method, which takes into account a larger range of the
FID in order to explain not only B0 inhomogeneities but relaxation as well
and which therefore allows for accurate field map estimation based on a
reference image and a single FID. Since the signal equation is non-linear,
the linear fitting of the original approach has to be replaced by a
non-linear optimization. The feasibility of the method is shown on in vivo
data.