Bruno Riemenschneider1, Jakob Assländer1, Pierre Levan1, and Jürgen Hennig1
We investigated a segmented version of the spherical stack-of-spirals
trajectory that retains highly efficient data sampling and signal recovery, but
grants more flexibility in data sampling compared to the single-shot version. Whole brain
acquisition with nominal isotropic resolutions of 3mm in
195ms and
2.25mm in 260ms using 3- and 4-fold segmentation have
been investigated. The faster read out along the slowest encoding direction
leads to reduced off-resonance artifacts in comparison to the single- shot
version, and higher sampling rates allow non-regularized reconstruction.