Abstract #2652
Initial Performance of the Diffusive Quantitative Imaging Phantom (DQIP): Thermal and SNR Characteristics Using a Clinical Protocol
Marc Buzzelli 1 , XiaoTian Li 2 , William Randazzo 1 , and Nathan Yanasak 1
1
Department of Radiology and Imaging, Medical
College of Georgia, Georgia Regents University, Augusta,
Georgia, United States,
2
Medical
College of Georgia, Georgia Regents University, Augusta,
Georgia, United States
Problem To characterize and apply thermal and
signal-to-noise corrections to DTI images of a
specialized quality assurance phantom via regression, to
facilitate future investigation of systematic biases in
clinical protocols. Methods A phantom was constructed
for diffusion imaging. Scans were obtained utilizing a
3T magnet, monitoring the temperature of the phantom
during the scans. FA and ADC maps were calculated.
Dependencies in temperature and SNR were determined, to
examine baseline systematic variability after
regression. Results/Conclusion FA and ADC show linear
dependence on temperature and an inverse dependence on
SNR. Intrinsic dependences of FA on SNR are significant
after regression.
This abstract and the presentation materials are available to members only;
a login is required.