The aim of this work is to design short and B0- and B1-robust inversion pulses by optimal control. A time-optimal control framework is used that incorporates variations within B0- and B1-fields. The optimized RF pulse is compared numerically to two hyperbolic-secant pulses and shows a very good efficiency over a broad set of B0-offsets and B1-scalings. Two phantom measurements are performed on a 3T MRI system for various scalings of B1 that verify the results, one with a cylindrical MRI phantom, the other one with oil and water with a contrast agent, demonstrating also the B0-robustness of the proposed pulse.