A Quadrature Birdcage/47Rx Coil Array for Acceleration Images on 3 T MRI

Jo Lee1,2, Sen Jia1,2, Liu Liu3, Xiaoliang Zhang4, and Ye Li1,2
1Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China, 2Shenzhen Key Laboratory for MRI, Shenzhen, China, 3United Imaging Healthcare, Shanghai, China, 4Department of Biomedical Engineering, State University of New York, Buffalo, NY, United States

We designed and built a quadrature-birdcage/ 47Rx head coil array for accelerated images on 3 T MRI. A 32-channel commercial head coil was used as a comparison. The inverse g-factor images and accelerated anatomical images both show that the 47-channel head coil has better acceleration ability. EPI images and cerebrovascular images further verified that the 47-channel head coil is capable of human’s brain studies.