The present study performed the repeatability and reproducibility test for T1, and T2 values estimated using MR fingerprinting (MRF) method. MRF data have been acquired over 30 days at three different centers (center-1 (USA): 1.5T and 3T, center-2 (USA): 3T and center-3 (Italy): 1.5T) using ISMRM/NIST MRI system phantom. MRF based estimated T1 and T2 values are compared with the vendor provided values and standard inversion recovery spin echo and a multiple single-echo spin-echo method as scanner Ground Truth (GT) values. The results show that the T1 and T2 values estimated using the MRF method are highly repeatable and reproducible.