To achieve dynamic and simultaneous access to R2* relaxation rates, B0 field inhomogeneities and water/fat separation, we developed a model-based reconstruction technique on BART for continuous acquisitions based on undersampled multi-echo multi-spoke radial FLASH. Beside spatial smoothness constraints on coil sensitivity and B0 field maps, L1 wavelet regularization is applied to the water, fat and R2* maps. Preliminary results of brain fMRI data demonstrate significant T2* change from 37 to 67 ms in the occipital visual cortex. In addition, R2* mapping of free-breathing liver with only 15 RF shots per frame (194 ms temporal resolution) reveals increased R2* during inspiration.