A recent novel approach to acquire high-resolution breast DWI uses a simultaneous multi-slice (SMS) SE-EPI sagittal acquisition. In EPI, Nyquist ghosts are typically corrected using a 3-line navigator, which often fails in SMS SE-EPI breast DWI due to low SNR, insufficient fat suppression, and larger B0 inhomogeneity. In this work we compare a referenceless ghost correction method, called Ghost/Object minimization, with the standard 3-line navigator in high-resolution breast DWI. Ghost/Object provides more reliable 1st-order ghost correction in a dynamic and slice-specific way, which improves image quality and reduces bias in ADC values compared to the standard correction.