The bright appearance of long-T2 tissues in DWI, termed “T2-shinethrough”, reduces the contrast between healthy tissue and cancer and is prominent in spin-echo based DWI acquisitions. In prostate DWI, the need to avoid T2-shinethrough has led to the acquisition of very high b-values in clinical practice, which may result in low SNR and other image artifacts. In this work, we have assessed the ability of stimulated-echo DWI to provide high contrast between PCa and healthy peripheral zone, without the need for high b-values. Preliminary results in 19 patients show reduced T2-shinethrough effects in stimulated-echo DWI compared with spin-echo DWI.
This abstract and the presentation materials are available to members only; a login is required.