A software tool (Multiparameter Analysis Method for B1 Acquisition - MAMBA) to design volume coils is presented for short T2* samples that optimizes relative SNR. MAMBA includes relevant RF coil parameters and evaluates performance. Sample and pulse sequence properties such as T2* relaxation times and echo time (TE) are considered. The tool is tested in a comparison of a commercial head coil and an optimized birdcage coil for imaging of an Egyptian mummy head.