In this work, a 32-channel transmit/receive (32Tx32Rx) head coil for 7T UHF MRI was developed and evaluated. The performance of this coil is compared to a commercially available 1Tx32Rx head coil regarding noise correlation, signal-to-noise ratio (SNR) and B1 homogeneity. The results indicate that high flexibility concerning RF shimming is provided by the 32Tx32Rx head coil that translates into a more homogeneous flip angle distribution compared to the 1Tx32Rx head coil. First 7T MR imaging results show the increased coverage of both the head and superior parts of the neck.