Based on the previous promising relaxation along fictitious field in nth frame (RAFFn) findings in myocardial infarct mouse model at 9.4 T, we adjusted RAFFn measurement to standard clinical 1.5 T scanner with γB1 = 500 Hz. We demonstrate with five myocardial infarct patients that fitted steady state from RAFF2 and RAFF3 weighted images can separate the infarct from remote myocardium using clinical main and RF fields. We also demonstrate association between RAFF3 steady state and extra cellular volume that is calculated based on pre and post contrast T1 maps and hematocrit.