High-field functional MRI of non-human primates provides elevated BOLD contrast in a well-established neuroscience model but requires an optimized coil design to boost the SNR, facilitate image acceleration while maintaining mechanical and electrical stability during the awake primate experiment. In this work we have constructed and tested 4-channel 9.4T and a 5-channel 3T receive-only array optimized for high-resolution whole-brain imaging of the anesthetized and awake marmosets.