A spin-echo EPI MR Elastography sequence was optimized so that multiple motion encoding gradient (MEG) cycles can be added to increase the motion sensitivity. Volunteer tests showed that comparing with the original one MEG version, optimized two or three MEG cycles provided higher Octahedral Shear Strain Signal-to-Noise Ratio (OSS-SNR), which means higher stiffness measurement precision. Global brain Images acquired with 2 and 3 MEG cycles are in most cases comparable in stiffness and OSS-SNR while images with 1 MEG tend to have a slightly lower OSS-SNR.