Meeting Banner
Abstract #1736

Ultra-low power transmitter for encoding non-MR signals in Magnetic Resonance (MR) recordings

Jan Raagaard Petersen1, Jan Ole Pedersen1,2, Vitaliy Zhurbenko1, Jan Henrik Ardenkjær-Larsen1, and Lars G. Hanson1,2

1Department of Electrical Engineering, Technical University of Denmark, Kgs. Lyngby, Denmark, 2Research Centre for Magnetic Resonance, Dept 714 Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark

Advancing Magnetic Resonance Imaging (MRI) technology requires integration of the MRI scanners with sensors and systems for monitoring various non-MRI signals. In this paper, we present design and integration of a low power AM radio transmitter into a 3T MRI scanner, which can be used for efficient collection of data from non-MRI sensors. The transmitter consumes only 1.3mW while transmitting 2.7µW at 120MHz with high frequency stability. The presented design is useful in low power applications requiring high frequency stability and is intended for wireless transmission of non-MR signal recordings during MRI scanning.

This abstract and the presentation materials are available to members only; a login is required.