Quantitation of MRSI data using water-scaling requires correction of the water signal for relaxation and CSF partial volume effects. We demonstrate the use of a rapid MRF sequence to characterize the water signal used to quantify MRS data, which we call WAter-scaling Quantification using MRF (WAQ-MRF) scan. WAQ-MRF provides subject-specific corrections of partial volume and relaxation effects for water-scaled data. By adding a one minute scan to a standard MRSI acquisition it is possible to eliminate the need for assuming literature values of relaxation and proton density to correct the water signal.
This abstract and the presentation materials are available to members only; a login is required.