To investigate the
optimal b-values of diffusion-weighted imaging (DWI), a prototype sequence with
integrated slice-specific shimming (iShim) was applied for visualizing thyroid malignant
nodules and making accurate differentiation between benign and malignant
nodules. DWI images were acquired with five b-values including 0, 800, 1000,
2000 and 3000 sec/mm2. The results show that the iShim-DWI with b-value of 800
s/mm2 can present the clearest delineation of malignant thyroid
nodules and that the images with b-value of 3000 s/mm2 had the best
performances in differentiating malignant from benign thyroid nodules.