High-resolution MRI of the lower jaw with an isotropic voxel size of 350 μm was performed at 3T using a 4-cm loop coil (LC) with and without an inductively coupled volume coil (ICC) enclosing the target region. Within this region, a much higher SNR was obtained and the resolution of anatomical details was strongly enhanced with the ICC. A new, custom-made holder for the LC improved the overall handling. Compared to earlier results, the scan time was reduced by 50%, now 2 minutes, to provide very high resolution MRI of 2-3 teeth.