Dixon Imaging with Golden Angle Stack of Stars Acquisition
Jan Hendrik Wlbern 1 , Mariya Doneva 1 , Holger Eggers 1 , Christian Stehning 1 , and Peter Brnert 1
1
Philips Research Europe, Hamburg, Hamburg,
Germany
Combined radial stack of stars k-space sampling with
dual-echo readout Dixon water-fat separation is
demonstrated. This is enabled by a radial phase
correction, which is performed on-the-fly without
requiring pre-scan calibrations relying on imaging data
only. Two types of radial sampling schemes are
considered: uniform angular sampling with alternating
readout directions and golden angle sampling. The phase
correction method preserves phase information for Dixon
methods, is robust to radial undersampling, stable over
long scan durations, and works for golden angle
acquisitions.