Abstract #3144
A 32 Channel Bi-lateral Breast Array for High Resolution Accelerated MR Imaging
R. O. Giaquinto 1,2 , R. G. Pratt 1 , W. M. Loew 1 , H. Friel 3 , L. Bickford 3 , C. Ireland 1 , B. Daniels 1 , B. Williams 1 , L. Haas 1 , J. M. Lanier 1 , K. M. Cecil 1,2 , M. Mahoney 2 , E. A. Morris 4 , and C. L. Dumoulin 1,2
1
Imaging Research Center, Cincinnati
Childrens Hospital Medical Center, Cincinnati, Ohio,
United States,
2
UC
College of Medicine, University of Cincinnati,
Cincinnati, Ohio, United States,
3
Philips
Healthcare, Best, Netherlands,
4
Memorial
Sloan Kettering Cancer Center, New York, New York,
United States
We have designed and fabricated a new 32 channel
ultra-high resolution 3T breast imaging coil array which
enables spatial resolution approaching that of X-ray.
The new coil permits acceleration factors of 16 or more
and full bi-lateral coverage of the breasts with an
isotropic resolution down to 0.3 mm. In this study,
benchmark comparisons were made with a commercially
available 7 channel breast array. The 32-channel coil
provided increased speed and resolution for DWI,
T1-weighted and T2-weighted images in phantoms and human
volunteers.
This abstract and the presentation materials are available to members only;
a login is required.
Join Here