Asymmetrically segmented loop phased coil for uniform RF field excitation at 7T
Seunghoon Ha 1 , Haoqin Zhu 1 , and Labros Petropoulos 1
1
R&D, IMRIS Inc., Minnetonka, MN, United
States
An asymmetric distributed capacitor capacitance loop
array structure that generated highly homogenous B1
field for 7T MRI was presented. The calculated B1 field
uniformity and coil sensitivity of the proposed coil was
superior to the identical size loop array with
symmetrically distributed capacitance as well as the
stripline array coil with the same length. The volume
shaped array loop coil assembly, consisting of the
proposed asymmetric distributed capacitance coil design
loops, was confirmed both through testing as well as
confirmed our approach through B1+ field simulation
study. Also, the proposed design will potentially help
the construction of larger size loop arrays to cover
more imaging volume at very high fields and obtain
uniform coverage and high sensitivity.