Abstract #3102
Evaluation on coupling strategies for ultra-high field MRI probe made of cylindrical dielectric resonator
Rui Liu 1 , Wei Luo 2 , Thomas Neuberger 3,4 , and Michael Lanagan 1,2
1
Engineering Science and Mechanics,
Pennsylvania State University, University Park,
Pennsylvania, United States,
2
Material
Research Insititute, Pennsylvania State University,
University Park, Pennsylvania, United States,
3
Huck
Institute of Life Science, Pennsylvania State
University, University Park, Pennsylvania, United
States,
4
Department
of Biomedical Engineering, Pennsylvania State
University, University Park, Pennsylvania, United States
In this study, different coupling methods for a
cylindrical dielectric resonator operating in TE01δ mode
were investigated and compared at 14T. The best coupling
method was selected based on the analysis of the
scattering parameters. It yielded a Q-factor around
716.2 and S21 value of -9.20 dB. The relationship
between different Q-factors and their effects on S21
were discussed. The results from this study will be
useful for the new designs on MRI probe head made of
cylindrical dielectric resonators.