Determination of Technically and Physiologically caused Variation of Parameters from DTI and BOLD MRI in Native Kidneys: A Repeatability Study
Maryam Seif 1 , Chris Boesch 1 , and Peter Vermathen 1
1
Depts. Radiology and Clinical Research,
University Bern, Bern, Switzerland
We evaluated the reproducibility of DTI and BOLD MRI
parameters in native kidneys of healthy volunteers by
performing three scans. Two scans were performed
back-to-back with one hour break in between, followed by
the third scan after three weeks. High reproducibility
was obtained for ADC, FA, and R2* from DTI and BOLD
measurements. The coefficients of variation between the
scans separated by 3 weeks were similar to those for the
back-to-back scans, indicating that the variances are
primarily due to technically caused variances between
scans rather than normal physiological variation over
time.