Abstract #0204
Optimized Combination of Magnetic Resonance Spectroscopy Signal from Multi-Element Coil Arrays
Liang Fang 1,2 , Minjie Wu 1 , Hengyu Ke 2 , Anand Kumar 1 , and Shaolin Yang 1,3
1
Department of Psychiatry, University of
Illinois at Chicago, Chicago, IL, United States,
2
School
of Electronic Information, Wuhan University, Wuhan,
Hubei, China,
3
Department
of Radiology, University of Illinois at Chicago,
Chicago, IL, United States
Multi-element coil arrays are commonly used in magnetic
resonance imaging (MRI) for parallel imaging. However,
how to combine the magnetic resonance spectroscopy (MRS)
signal from a multi-element receiving coil array has
been much less studied than imaging. In this study, we
proposed a new method for theoretical optimal
combination no matter the noise is correlated or
uncorrelated among different coil elements. The
validation experiments were performed with different
voxel locations/sizes on a 3T MRI scanner with an
8-element head coil. The results verified the superior
performance of the proposed method compared with the
recently published and other routine combination
methods.
This abstract and the presentation materials are available to members only;
a login is required.
Join Here