Interplatform Variability of Liver and Spleen MR Elastography
Temel Kaya Yasar 1 , Octavia Bane 1 , Cecilia Besa 1 , Stephan Kannengiesser 2 , and Bachir Taouli 1
1
Icahn School of Medicine at Mount Sinai, New
York, NY, United States,
2
Siemens
Healthcare, Germany
There is insufficient knowledge on variability of MR
Elastography (MRE) between different platforms. In this
study, 6 underwent MRE on 1.5T Siemens and 3.0T GE
systems using a 2D-GRE sequence. Liver and spleen
stiffness values obtained from the 2 scanners were
compared. Excellent concordance was observed between
scanners (CV of 9.8% and 12.0% for liver and spleen,
respectively). It remains to be seen whether different
sequences on different platforms will provide similar
results.