Abstract #2456
Intrinsic motion correction for radial cardiac T 2 mapping through alternating T 2 preparation duration
Helene Feliciano 1,2 , Matthias Stuber 1,2 , and Ruud B. van Heeswijk 1,2
1
Radiology, University Hospital (CHUV), and
University of Lausanne (UNIL), Lausanne, Switzerland,
2
Center
for Biomedical Imaging (CIBM), Lausanne, Switzerland
A radial T
2
mapping
sequence that makes use of a cyclically alternating T
2
Prep
was developed at 3T to avoid the adverse effects of
respiratory or RR variability on T
2
quantification.
Bloch simulations and phantom experiments were performed
to compare the thus obtained T
2
maps
with that of a standard sequential T
2
mapping
approach. In 9 healthy volunteers it was confirmed that
more robust T
2
mapping
is obtained with alternating T
2
preparation.
This abstract and the presentation materials are available to members only;