A RF supervision concept that considers local SAR is an essential pre-requisite for high performance pTX applications. It needs to provide at least one online RF supervision component that detects violation of local SAR limits in real-time. Using the MR signal processing unit for that fails to match real-time requirements. In this study, an online local SAR calculation was implemented on the image reconstruction PC of a MRI system. To overcome latency issues, a real-time logic in conjunction with a real-time global SAR supervision unit is used to guarantee patient safety at all times