Meeting Banner
Abstract #4327

Real-Time Image Guided Targeting of MRI Compatible Robotic Assisted Breast Biopsy & Therapeutic System

Steven Roys1, Bo Yang2, Mathew Philip3, Jaydev Desai2, Rao Gaullapalli1

1Diagnostic Radiology and Nuclear Medicine, University of Maryland School of Medicine, Baltimore, MD, United States; 2Mechanical Engineering, University of Maryland, College Park, College Park, MD, United States; 3Robin Medical, Inc., Baltimore, MD, United States


MR guided robotic assisted biopsy and therapeutic systems can be very useful as they can take advantage of the excellent soft tissue contrast especially to reach deep seated tumors that are difficult to target accurately. In this study we have evaluated a master-slave robotic system designed for breast biopsy and interventions that incorporates a position sensing device. The accuracy of the targeting under continuous imaging and the time taken to target deep seated tissues using a needle were measured. Accuracy of targeting was within 2mm and the targeting procedure time was ~20minutes.