1Institute
for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland; 2Skope
Magnetic Resonance Technologies , Zurich, Switzerland
Gradient-based 3D positioning of an NMR field probe relies on independent information on all 3 gradient axes. Position encoding using sinusoidal gradient tones in the kHz range suffers from inductive coupling among 0th and 1st order fields. Field decoupling is demonstrated, the improvement is quantified (accuracy < 100 &[micro]m ) and the method is applied to in-vivo real-time prospective motion correction in the brain.