The goal of this study was to analyze the effect of controlled compression and tension on the dynamic shear modulus for the two directions of wave propagation orthogonal to the particle motion. These mediums were a 100% PVC phantom and an ex vivo specimen of bovine liver. For both the phantom and the specimen, the measured shear modulus in both directions increases with the absolute value of the axial strain under both compression and tension. This data provide motivation to further investigate the effects of pre-strain and their clinical utility to improve the information obtained from MRE.