Grace Sooyeon Kim1, Daniel Burje Chonde1,
Thomas Benner1, Michael Hamm2, Alma Gregory Sorensen1,
Ciprian Catana1
1A.A.
Bone/air segmentation is a significant challenge in developing the MRI-based attenuation correction (AC) method. We determined good reproducibility of a dual-echo ultra-short echo time (DUTE) sequence-based AC method used in ten brain tumor patients over four time points (relative SD between 2.17 to 13.40%). We found that the DUTE-based method yielded a 85.30% average agreement with respect to the silver-standard segmented CT-based map. The greatest challenges with this method involve segmentation of the full thickness of the skull and distinguishing air from tissue in difficult sinus regions. Further development of the DUTE-based AC method is under way.