Meeting Banner
Abstract #3793

Bi-Planar Shim Coil Designed by Stream Function Method Improves B0 Homogeneity Along Z-Axis

Daiki Tamada1, Yasuhiko Terada1, Katsumi Kose1

1Institute of Applied Physics, University of Tsukuba, Tsukuba, Ibaraki, Japan


A new method using stream function was proposed to correct inhomogeneity of the magnetic field with single-channel shim coils. The inhomogeneity achieved by the stream function method was comparable to that achieved by the target field method. Moreover, the inhomogeneity along z-axis was also corrected, which could not be done by the target field method. The method presented here provides large three dimensional homogeneity without multi-channel shim coils, which leads to the enlargement of available space for samples.