Meeting Banner
Abstract #2901

Adapted Tx-SENSE Excitation to Account for Inhomogeneous Slice Refocusing at 7T

Tomasz Dawid Lindel1,2, Frank Seifert1,2, Martin Dietterle1,2, Thoralf Niendorf2, Bernd Ittermann1,2

1Physikalisch-Technische Bundesanstalt, Braunschweig & Berlin, Germany; 2Berlin Ultrahigh Field Facility, Max-Delbrck-Centrum fr Molekulare Medizin, Berlin, Germany


2D spatial excitation using Tx-SENSE is frequently combined with a conventional, slice selective refocusing pulse to restrict the signal in the third dimension. Severe B1+ inhomogeneities of the refocusing pulse at 7T compromise the resulting image quality even if B1 shimming is applied. We present phantom experiments with adapted Tx-SENSE pulses to address this problem.