Failure to achieve the CPMG conditions for a 3D FSE sequence results in significant image artifacts, particularly for off-isocenter imaging due to system non-idealities. We prospectively correct the sequence by measuring phase errors in the x and z directions during prescan and update the sequence parameters so that the modified sequence meets the CPMG conditions. This iterative correction allows us to collect images without banding artifacts.