PatLoc imaging with non-linear quadrupolar gradients has been developed and realised for rapid MR imaging. This gradient coil is driven within a clinical 3T scanner with 6 gradient channels for simultaneous use of the linear and PatLoc gradients. For rapid imaging the gradient coil performance and accuracy is especially important. This abstract describes the measurement procedure and the compensation of eddy currents from around 300Hz to 30Hz, which improves image quality significantly.