Christopher Chad Quarles1, John Christopher
Gore1, Thomas Edison Yankeelov1
1Radiology &
Radiological Sciences,
We have previously demonstrated that dual-echo DSC-MRI can be used to separate and quantify T1 and T2* based contrast agent leakage effects and, when combined with a pre-contrast T1 map, quantify the DCE-MRI parameters, Ktrans and ve. To further verify this approach we now compare DSC-MRI and DCE-MRI derived Ktrans and ve values in two brain tumor models with dissimilar perfusion characteristics and contrast agent dynamics. The DSC-MRI and DCE-MRI parameters were highly correlated and in good agreement as no significant differences were found in either of the tumor models.