florence
franconi1, Jean-Jacques Le Jeune2, Pascal Richomme1,
Laurent Lemaire2
1PIAM, Universit d'Angers, Angers,
France, Metropolitan; 2UMR-S646, INSERM, Angers, France
SIgnal
Response MApping to dephaser (SIRMA) method is proposed to quantify
susceptibility gradient. In gradient echo acquisitions, the SIRMA method
measures the echo shifts in k-space of susceptibility affected spins from a
series of dephased images collected with additional incremental slice
refocusing gradient offset or incremental reconstruction window
off-centering. SIRMA applicability and performances have been demonstrated in
vitro through quantization of susceptibility gradient induced in a cylinder
model and in vivo through the quantitative detection of SPIO distribution
volume. With respect to its quantitative nature, its computational
simplicity, this method deserves further attention for application in
molecular or cellular imaging.