Bruno Quesson1, Mathilde Merle1,
Max Khler2, Charles Mougenot3, Sebastien Roujol1,
Baudouin Denis de Senneville1, Chrit T. Moonen1
1laboratory for molecular and
functional imaging, bordeaux, France; 2Philips healthcare, Vantaa,
Finland; 3Philips Healthcare, bordeaux, France
The
partial obstruction of the High Intensity Focused Ultrasound beam by the rib
cage complicates the treatment of liver tumors. A method for selective
deactivation of the transducer elements located in front of the ribs
(visualized on 3D anatomical MR images) is proposed. The effectiveness of the
method for HIFU liver ablations is demonstrated ex vivo and in vivo in pigs
during breathing with real-time, motion compensated, MR thermometry. No loss
in heating efficacy was observed at the focal point and an important
reduction of the heating in tissues surrounding the bones was obtained with
deactivation of the transducer elements.