Hector Sanchez1, Michael Poole1,
Adnan Trakic1, Stuart Crozier1
1Research Group/Affiliation: EMI,
School of Information Technology & electrical Engineering, Brisbane, QLD,
Australia
MRI
requires rapidly switched magnetic field gradients. This time-dependent
magnetic fields induce eddy currents in nearby conducting structures. These
currents generate detrimental transient magnetic fields in the region of
interest (ROI) and hence, current compensation is required to minimize the
consequential image distortion. In order to apply successfully current
compensation techniques, it is required that the primary and the secondary
magnetic fields possess a similar spatial form in the ROI. In this work we
present two approaches for gradient coil design that produces gradient fields
with characteristics similar to those produced by the eddy currents.